Aniline Detection Tubes 5/a (1 - 20 ppm)
Draeger Short-term Tubes have proven to be a very cost-effective and reliable method for the measurement of gases.
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapours can be detected and measured, e.g.: Determining and measuring concentration peaks, Measuring personal exposure in the inhalation area, Detecting leaks in pipelines, gas installations etc., and Air analysis in sewers, shafts and confined spaces.
Aniline Detection Tubes 5/a (1 - 20 ppm)
Standard Measuring Range: |
1 to 20 ppm |
Number of Strokes (n): |
5 to 25 |
Time for Measurement: |
max. 3 min. |
Standard Deviation: |
±30 % |
Color Change: |
white ?> red |
Temperature: |
15 to 40° C |
Absolute Humidity: |
< 50 mg H2O / L |
Reaction Principle: |
Aniline + Furfurol ?> Dianiline derivative of Hydroxyglutacondialdehyde |
Cross Sensitivity: |
N,N-Dimethylaniline is not indicated.
Ammonia concentrations up to 50 ppm do not affect the indication, higher ammonia concentrations will cause plus errors. |
Additional Information |
Before performing the measurement the reagent ampoule must be broken. The granular contents must be shaken out of the ampoule by tapping the side of the tube. During the measurement. the Draeger Tube must be held vertically with the inlet of the tube up. |