Natural Gas Test Detection Tubes - Qualitative
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapors can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces
Natural Gas Test Detection Tubes - Qualitative
Standard Measuring Range: |
Qualitative determination of natural gas |
Number of strokes n: |
5 |
Time for Measurement: |
approx. 100 sec |
Standard Deviation: |
50 % |
Color Change: |
white ??> brown green to gray violet |
Temperature: |
0 to 50 °C |
Absolute Humidity: |
max. 40 mg H2O / L |
Reaction Principle: |
a) CH4 + KMnO4 + H2S2O7 ??> CO b) CO + I2SO5 ??> I2 + CO2 |
Cross Sensitivity: |
Due to the reaction principle a number of other organic compounds such as propane and butane are indicated as well. Carbon monoxide is also indicated. It is not possible to differentiate between different compounds.
|