Nitrogen Dioxide Detection Tubes 2/c (2 - 100 ppm)
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapors can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces
Nitrogen Dioxide Detection Tubes 2/c (2 - 100 ppm)
Standard Measuring Range: |
5 to 100 ppm / 2 to 50 ppm |
Number of strokes n: |
5 / 10 |
Time for Measurement: |
approx. 1 min / approx. 2 min |
Standard Deviation: |
± 10 to 15 % |
Color Change: |
yellow green ??> blue gray |
Temperature: |
10 to 40 °C |
Absolute Humidity: |
max. 30 mg H2O / L |
Reaction Principle: |
NO2 + diphenylbenzidine ??> blue gray reaction product |
Cross Sensitivity: |
Within their TLV, ozone or chlorine do not interfere with the reading. Higher concentrations are indicated, however with differing sensitivity period. Nitrogen monoxide is not indicated. |
Nitrogen Dioxide Detection Tubes 2/c (2 - 100 ppm)
There are no reviews yet.