Toluene Detection Tubes 50/a (50 - 400 ppm)
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapours can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces
Toluene Detection Tubes 50/a (50 - 400 ppm)
Standard Measuring Range: |
50 to 400 ppm |
Number of strokes n: |
5 |
Time for Measurement: |
approx. 1.5 min |
Standard Deviation: |
± 10 to 15 % |
Color Change: |
white ??> brown |
Temperature: |
0 to 30 °C |
Absolute Humidity: |
5 to 12 mg H2O / L |
Reaction Principle: |
Toluene + I2O5 + H2SO4 ??> I2 |
Cross Sensitivity: |
Xylenes will be indicated with different sensivitites. Benzene causes a total discoloration into yellow. Petroleum hydrocarbons cause a total discoloration into a diffuse red brown. Methanol, ethanol, acetone and ethyl acetate do not disturb the discoloration in the range of TLV-values. |