Perchloroethylene Detection Tubes 2/a (2 - 300 ppm)
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapors can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces
Perchloroethylene Detection Tubes 2/a (2 - 300 ppm)
Standard Measuring Range: |
20 to 300 ppm / 2 to 40 ppm |
Number of strokes n: |
1 / 5 |
Time for Measurement: |
approx. 30 sec / approx. 3 min |
Standard Deviation: |
15 to 20 % |
Color Change: |
slightly greenish ??> grey blue |
Temperature: |
15 to 30 °C |
Absolute Humidity: |
< 25 mg H2O / L |
Reaction Principle: |
a) CCl2=CCl2 + MnO4- ??> Cl2 b) Cl2 + Diphenylbenzidine ??> gray blue reaction product At higher concentrations there may be created a reddish area at the beginning of the indicating layer. |
Cross Sensitivity: |
Other chlorinated hydrocarbons, free halogens and hydrogen halides are also indicated. Petroleum hydrocarbon vapors result in a shortened reading once they exceed the following concentrations: 50 ppm with 5 strokes or 500 ppm with 1 strokes |
Perchloroethylene Detection Tubes 2/a (2 - 300 ppm)
There are no reviews yet.