Perchloroethylene Detection Tubes 0.1/a (0.1 - 4 ppm)
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapors can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces
Perchloroethylene Detection Tubes 0.1/a (0.1 - 4 ppm)
Standard Measuring Range: |
0.5 to 4 ppm / 0.1 to 1 ppm |
Number of strokes n: |
3 / 9 |
Time for Measurement: |
approx. 3 min / approx. 9 min |
Standard Deviation: |
± 15 to 20 % |
Color Change: |
light gray ??> blue |
Temperature: |
15 to 30 °C |
Absolute Humidity: |
max. 30 mg H2O / L |
Reaction Principle: |
a) CCl2=CCl2 + MnO4- ??> Cl2 b) Cl2 + diphenylbenzidine ??> blue reaction product |
Cross Sensitivity: |
Other chlorinated hydrocarbons, free halogens and hydrogen halides are also indicated. Petroleum hydrocarbon vapors result in a shortened reading once they exceed the following concentrations: 40 ppm with 9 strokes or 160 ppm with 3 strokes |