Sulfur Dioxide Detection Tubes 0.5/a (0.5 - 25 ppm)
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapors can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces
Sulfur Dioxide Detection Tubes 0.5/a (0.5 - 25 ppm)
Standard Measuring Range: |
1 to 25 ppm / 0.5 to 1 ppm |
Number of strokes n: |
10 / 20 |
Time for Measurement: |
approx. 3 min / approx. 6 min |
Standard Deviation: |
± 10 to 15 % |
Color Change: |
gray blue ??> white |
Temperature: |
15 to 30 °C |
Absolute Humidity: |
max 20 mg H2O / L |
Reaction Principle: |
SO2 + I2 + 2 H2O ??> H2SO4 + 2HI |
Cross Sensitivity: |
Measurement is not possible in the presence of H2S. Nitrogen dioxide will shorten the reading. |