Phosphine Detection Tubes 50/a (15 - 3000 ppm)
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapors can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces
Phosphine Detection Tubes 50/a (15 - 3000 ppm)
Standard Measuring Range: |
50 to 1,000 ppm |
Number of strokes n: |
3 |
Time for Measurement: |
approx. 2 min |
Standard Deviation: |
± 10 to 15 % |
Color Change: |
yellow ??> brown black |
Temperature: |
0 to 50 °C |
Absolute Humidity: |
< 40 mg H2O / L |
Reaction Principle: |
PH3 + Au3+ ??> Au (colloidal) |
Cross Sensitivity: |
Arsine and antimony hydride are indicated, but with different sensitivities.
Hydrogen sulfide, mercaptans, ammonia, carbon monoxide, sulfur dioxide and hydrochloric acid in the TLV range do not
interfere. |
Extension of the Measuring Range: |
Using n = 10, multiply the reading by 0.3; the range of measurement is 15 to 300 ppm. |