Phosgene Detection Tubes 0.02/a (0.02 - 1 ppm)
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapors can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces
Phosgene Detection Tubes 0.02/a (0.02 - 1 ppm)
Standard Measuring Range: |
0.02 to 1 ppm / 0.02 to 0.6 ppm |
Number of strokes n: |
20 / 40 |
Time for Measurement: |
approx. 6 min / approx. 12 min |
Standard Deviation: |
± 10 to 15 % |
Color Change: |
white ??> red |
Temperature: |
0 to 40 °C |
Absolute Humidity: |
3 to 15 mg H2O / L |
Reaction Principle: |
COCl2 + Aromatic amine ??> red reaction product |
Cross Sensitivity: |
Chlorine and hydrochloric acid lead to plus errors and, at higher concentrations, to bleaching of the indicating layer. Concentrations of phosgene above 30 ppm will also lead to bleaching of the indicating layer. |
Additional Information: |
High concentrations of phosgene will be not indicated! |