Gasoline Detection Tubes (30 to 2000 ppm)
Measures Gasoline with a Range of 30 to 2000 ppm.
- Measuring Range: 30 to 1000 ppm, 1000 to 2000 ppm
- Number of Pump Strokes: 2, 1
- Correction Factor: 1, 2
- Sampling Time: 2 minutes per pump stroke
- Detecting Limit: 5 ppm (2 pump strokes)
- Changes from Yellowish brown to Greenish brown
- Reaction Principle: CnHm + Cr6+ + H2SO4 → Cr3+
- Relative Standard Deviation: 10% (for 30 to 300 ppm), 5% (for 300 to 1000 ppm)
- Corrections for temperature & humidity: Unnecessary
- 3 Year Shelf Life
- Store the tubes at cool and dark place.
Possible coexisting substances and their interferences:
Substance |
Interference |
Changes color by itself to |
Alcohols |
+ |
Greenish brown |
Aromatic hydrocarbons |
+ |
Greenish brown |
Esters, Ethers |
+ |
Greenish brown |
Organic solvents (≧ C3) |
+ |
Greenish brown |
Sulfur dioxide |
+ |
Greenish brown |
Acetylene |
+ |
Blackish brown |
Hydrogen sulfide |
+ |
Blackish brown |
Other substance measurable with this detector tube:
Substance:
|
Correction Factor:
|
No. of Pump Strokes:
|
Measuring Range (ppm):
|
Allyl Chloride |
Factor: 34 |
1/2 |
0.1 to 3.4% |
Heptane |
Factor: 2 Factor: 1 |
1 2 |
100 to 2000 30 to 1000 |
Isobutene |
Factor: 22 |
1 |
0.07 to 2.2% |