Gastec Detection Tube No. 163L measures Ethylene Oxide with a Range of 0.4 to 350 ppm.
- Measuring Range:
- 0.4 to 1 ppm
- 1 to 100 ppm
- 100 to 350 ppm
- Number of Pump Strokes: 4, 2, 1
- Correction Factors: 0.4, 1, 3.5
- Sampling Time: 3 minutes per pump stroke
- Detecting Limit: 0.1 ppm (4 pump strokes)
- Changes from Yellow to Reddish Brown
- Corrections for temperature & humidity: Temperature correction is necessary
- Reaction Principle:
- C2H4O → 2HCHO
- 3HCHO + (NH2OH)3H3PO4 → H3PO4
- H3PO4 + Base → Phosphate
- Relative Standard Deviation: 15% (for 1 to 20 ppm), 10% (for 20 to 100 ppm)
- 1 Year Shelf Life (in the refrigerator)
- Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
- Comes in a Box of 5 pre tubes and 5 detector tubes
Possible coexisting substances and their interference:
Substance |
Interference |
Changes color by itself to |
Aldehydes |
+ |
Reddish brown |
Ketones |
+ |
Reddish brown |
Other substance measurable with this detector tube:
Substance: |
Correction Factor: |
No. of Pump Strokes: |
Measuring Range: |
Epichlorohydrin |
1.2 |
2 |
1.2 to 120 ppm |
Propylene oxide |
1.0 |
1 |
1 to 100 ppm |