Hydrogen Sulfide (H2S) Detection Tubes (1 to 40 ppm)
Gastec Detection Tube No. 4LK measures Hydrogen Sulfide with a Range of 1 to 40 ppm.
- Measuring Range:
- 1 to 2 ppm
- 2 to 20 ppm
- 20 to 40 ppm
- Number of Pump Strokes: 2, 1, 1/2
- Correction Factor: 1/2, 1, 2
- Sampling Time: 1 minute per pump stroke
- Detecting Limit: 0.25 ppm (2 pump strokes)
- Changes from White to Brown
- Corrections for temperature & humidity: Unnecessary
- Reaction Principle: H2S + Pb(CH3COO)2 → PbS + 2CH3COOH
- Relative Standard Deviation: 10% (for 2 to 6 ppm), 5% (for 6 to 20 ppm)
- 3 Year Shelf Life
- Store the tubes at cool and dark place.
Possible coexisting substances and their interferences:
Substance |
Concentration |
Interference |
Changes color by itself to |
Ozone |
≧ 10 ppm |
Bleaching |
No |
Sulfur dioxide |
≧ 1/2 |
+ |
No |
Nitrogen dioxide |
≧ 1/4 |
- |
No |
Mercaptans |
|
No |
No |
Hydrogen Sulfide (H2S) Detection Tubes (1 to 40 ppm)
There are no reviews yet.