Methanol Detection Tubes (2 to 56 ppm)
Gastec Detection Tube No. 111LL measures Methanol with a Range of 2 to 56 ppm.
- Number of Pump Strokes: 4, 2
- Correction Factor: 1, 2.8
- Sampling Time: 2 minutes per pump stroke
- Detecting Limit: 0.2 ppm (4 pump strokes)
- Changes from Pale Yellow to Pale Bluish Green
- Corrections for temperature & humidity: Temperature correction is necessary
- Reaction Principle: CH3OH + Cr6+ + H3PO4 → Cr3+
- Relative Standard Deviation: 15% (for 2 to 5 ppm), 10% (for 5 to 20 ppm)
- 2 Year Shelf Life
- Store the tubes at cool and dark place.
Possible coexisting substances and their interferences:
Substance |
Interference |
Changes color by itself to |
Alcohols |
+ |
Pale bluish green |
Aromatic hydrocarbons |
+ |
Pale bluish green |