Phosphine (PH3) Detection Tubes (0.05 to 9.8 ppm)
Gastec Detection Tube No. 7LA measures Phosphine with a Range of 0.05 to 9.8 ppm.
- Measuring Range:
- 0.05 to 0.1 ppm
- (0.1) to 1.5 ppm
- 1.5 to 2.5 ppm
- 2.5 to 9.8 ppm
- Number of Pump Strokes: 10, 5, 3, 1
- Correction Factor: 1/2, 1, 1.7, 6.5
- Sampling Time: 1.5 minutes per pump stroke
- Detecting Limit: 0.02 ppm (10 pump strokes)
- Changes from Yellow to Red
- Reaction Principle:
- PH3 + 3HgCl2 → P(HgCl)3 + 3HCl
- HCl + Base → Chloride
- Relative Standard Deviation: 10% (for 0.1 to 0.4 ppm), 5% (for 0.4 to 1.5 ppm)
- 2 Year Shelf Life
- Store the tubes at cool and dark place.
Possible coexisting substances and their interferences:
Substance |
Concentration |
Interference |
Changes color by itself to |
Arsine |
≧ 1/10 |
+ |
Red |
Diborane |
|
No |
No |
Hydrogen chloride |
|
No |
No |
Phosphine (PH3) Detection Tubes (0.05 to 9.8 ppm)
There are no reviews yet.