Methyl Bromide Detection Tubes 0.1/a (0.1 - 50 ppm)
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapors can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces
Methyl Bromide Detection Tubes 0.1/a (0.1 - 50 ppm)
Standard Measuring Range: |
0.1 to 5 ppm / 5 to 50 ppm |
Number of Strokes (n): |
10 / 2 |
Time for Measurement: |
approx. 8 min / approx. 4 min. |
Standard Deviation: |
15 to 20 % |
Color Change: |
bright -> green |
Temperature: |
2 to 40 C |
Absolute Humidity: |
< 40 H2O / L |
Reaction Principle: |
a) CH3Br + H2S2O7 + chromate -> Br2 c) Br2 + o-tolidine -> green reaction product |
Cross Sensitivity: |
Carbon tetrachloride: < 2 ppm no reading
In the presence of perchloroethylene or trichloroethylene, a methyl bromide measurement is not possible!
Sulfuryl fluoride, phosphine, ethylene oxide, ammonia, hydrogen cyanide, chloropicrin and formaldehyde are not indicated below their threshold values.
2 ppm ethylene dibromide is indicated with approximately the same sensitivity.
0.5 ppm vinyl chloride is indicated with a reading of less than 0.1 ppm.
2 ppm 1.1 dichloroethylene is not indicated and 20 ppm
1.2 dichloroethylene is indicated with a low reading of 3 ppm. |