When calibrating or performing a bump test, it is important to use a regulator with flow rate that falls within the manufacturer's recommendations. Pumped instruments may generally be used with a demand flow regulator (DFR), which is the most accurate and efficient way to apply gas to the sensors.
Please consult your gas detector manual for more information on your specific gas monitor and bump test and calibration procedures.
Draeger |
PAC 5000, 7000 |
0.5 LPM |
|
X-AM 2500 |
0.5 LPM |
Honeywell Analytics (BW) |
GasAlertMicroClip |
0.5 LPM |
|
GasAlertMax XT II |
0.5 LPM, DFR |
|
GasAlert Micro 5 |
0.5 LPM |
MSA |
Altair 4X |
0.25 LPM |
|
Altair 5X |
DFR |
RAE Systems |
ToxiRAE Pro |
0.5 to 1.0 LPM |
|
QRAE 3 |
0.5 to 1.0, DFR |
|
MicroRAE |
0.5 to 1.0 LPM |
|
MultiRAE |
0.5 to 1.0 LPM, DFR |
RKI Instruments |
GX-2009 |
0.5 LPM |
|
GX-6000 |
DFR |
- DFR = Demand Flow Regulator
- LPM = Liters Per Minute
* For reference purposes only. Consult gas detector manual for exact manufacturer recommendations.